ISO 12714:1999 (R2023)
Non-destructive testing — Acoustic emission inspection — Secondary calibration of acoustic emission sensors
Summary
This International Standard specifies a method for the secondary calibration of acoustic emission sensors as receivers of elastic waves at the surface of a solid medium. The International Standard is applicable to laboratory tertiary standard sensors and acoustic emission applications sensors. The secondary calibration yields the frequency response of a sensor to waves of a type normally encountered in acoustic emission work, namely Rayleigh waves. The source producing the signal used for the calibration is mounted on the same surface of a test block as the sensor under test (SUT). The sensitivity of the sensor is determined for excitation within the range of 100 kHz to 1 MHz. Sensitivity values are usually determined at frequencies approximately 10 kHz apart. The units of the calibration are volts per unit of mechanical input (displacement, velocity or acceleration).
Notes
90.93 : Norme internationale confirmée
Technical characteristics
| Publisher | International Organization for Standardization (ISO) |
| Publication Date | 07/15/1999 |
| Confirmation Date | 07/13/2023 |
| Edition | 1 |
| Page Count | 16 |
| EAN | --- |
| ISBN | --- |
| Weight (in grams) | --- |
| Brochures |
|
No products.
Previous versions
15/07/1999
Active
, Confirmed
Most Recent